The standard MHJ-IIIB1 includes a pneumatic-controlled and pneumatic-driven anaethesia ventilator SC-M 3A 1, an anaesthesia circuit model IV and an economic anaestheisa vaporizer DZFG. MHJ-III B1 is suitable for use in various hospital for closed-anaesthesia,semi-closed anaesthesia and semi-open anesthesia,and for oxygen inhalation during resuscitation. |
| A flowmeter with ORC system ensures oxygen concentration not less than 25%. |
| With automatic N2O cut-off |
| O2deficiency alarm |
| SC-M 3A 1 anesthesia ventilator,with easy operation and low air consumption |
| Top plate intended for all kinds of monitors |
Main technical specifications |
Supply pressure:0.27~0.55MPa
|
Flow control:
O2 0.1~ 10L /min
;
N2O 0.1~ 10L /min
|
.ORC system: Oxygen not less than 25% within mixture gas
|
| N2O automatic cut-off:Active when oxygen pressure less than 0.1MPa
|
| O2 flush:50± 10L /min at O2 pressure of 0.45MPa |
| .Adjustment concentartion of Enflurane Vaporizer:0~5Vol%
|
| Airway pressure limit range(APL): 0.5~7kpa |
SC-M 3A 1 anaesthesia ventilator working type: Pneumatic-controlled and pneumatic-criven, time cycle, volume constant
|
Ventilation frequency:6~ 40L /min
|
Minute volume:3~ 18L /min |
| I:E ration:1:1.5 (fixed)
|
| Gas consumption:less than 0.8L /min(for control part)
|
| .
Dimensions:516×600× 1370MM (L×W×H) |
| Weight: 60kg |
| No need of power supply. Only compressed oxygen for pneumatic driving and pneumatic control. |
|