It adopts a two-stage pressure regulator and a high precision dual-tube flowmeter to provide an accurate fresh gas output.Its build-in safety devices such as ORC system, oxygen deficiency alarm and N20 cut-off make anaesthesia more safe. The specially constructed inter-lock mechanism may guarantee only one of the two parallel vaporizers to be switched on at a time.The selected vaporizers from world famous companies as Drager 19.3,
Ohmeda TEC-5 and Penlon Sigma Delta satisfy different requirements of customers.
With MHJ -1C it is possible to act low-flow anaesthesia because of its unique integrated circle
system-a system that keeps the inhaled tidal volume from fresh gas. So it not only saves anaesthesia
agent but also reduces pollution to operation room. |
Main Specifications: |
| .Flow:O2 0.1~ 15L /min
;
N2O 0.05~ 10L /min
;
ORC O2 = 25% |
| O2 Flush: 35~ 75L /min |
Oxygen deficiency alarm and N2O cut-off :when O2 = 0.2Mpa,audible alarm occurs and lasts for at least 7 seconds ;
when O2 = 0.14Mpa,N2O flow decreases correspondingly ;
when O2 = 0.04Mpa,N2O flow is cut off |
| Integrated breathing circle system:
APL adjustment range:0.5~7kPa
;
CO2 absorber acpcity: 1.6L
|
Vaporizer: |
Ambient temperature 15~ 35 ? |
Input flow 0.25~ 15L /min |
Output back pressure:-10kPa~20kPa |
| Output concentration:0.2~5Vol%±20%(OR±0.2%) |
Anesthesia Ventilator: |
A microprocessor-controlled ventilator,which is time-cycled and constant-volume,has a clear digital reading for all set breathing parameters.Because it is electric-driven and electric-controlled,the ventilator saves large amount of oxygen that is usually used only as driving gas. |
Specifications:
Frequency:6-60bmp ;Tidal volume:100~1100mL ;
I:E ratio 1:1~1:9.9 adjustable
|
Monitor: |
| Respiratory monitor applies microprocessor to monitor Tidal Volume,Frequency,MV airway pressure,and inhaled oxygen concentration.It has real time curve display and a visual-and-audible alarms.A Chinese menu and single-key operation is userfriendly. A high-definition large screen with wide vision angle make all-directional inspection available. |
| Specification:
Frequency:6-60bmp
O2 measuring range 0~99%
Tidal Volume:100~1500mL
Minute volume:0 -40L /min
Airway pressuer:0-9.9kPa |
|