MHJ- IIIB2 multifunction anaesthesia machine

Back to main menu

 
 
 
 
 
 
Standard parts of MHJ-IIIB2 include anaesthesia ventilator SC-M 3A ,a compact vaporizer and airway
gauge.It is possible to be furnished with other types of vaporizers and electronic monitorings to
make an ideal multifunction anaesthesia machines for modern hospital.

Feratures:

ISOcircuit system adaptable to various monitoring

New type of flowmeter ORC system (O2concentration no less than 25%),N2O cut off safety unit.

Conpact Enflurane Vaporizer:Vaporized drug comcentration adjustable

The SC-M 3A pneumatic anaesthesia ventilator was comparable small consumption of gas(one cylinder of Oxygen for six hours of operation),and precise parameters for control

Airway pressure gauge for real-time display of respiration of patient

Technical Data:

Flow control range:O2 ( 0.1 ~ 10L /min )

Gas Supply:0.27 ~ 0.55MPa

O2 Flush:35 ~ 70L /min

N2O automatic cut-off:active when O2 pressure lower than 0.1MPa

Non-compensation Vaporizer(Enflurane), Concentration adjustable(0 ~ 5%)

Anaesthesia Ventilator: Control type:Pneumatically driven and controlled,time cycle and volume constant:
                                   Gas supply:0.27~0.55Mpa
                                   Frequency:6~40cycles/min
 
 