Home        Company          Products        Contact us
   
       
     

100mA Medical X-ray Camera

Back to main menu

 
 
 
 
 
 
 

MAIN CHARACTERISTICS

 

1.Single bed and single bulb.

2.Design according to the electric mechanism of middle-size machine.The function is full and reasonable.

3.The high-tension system is of march-past structure.Equipped with afullwave rectification high-voltage generator and independent X-ray tube unit,with large output power and reliable and stable operation.

4.Supply voltage V,roentgenoscopy,photograpy KV and stepless adjustable.

5.Equipped with the manostat for filament of X-ray tube and the spacecharge compensator as to ensure the stable and accurate output.

6.Equipped with the capacity limitation and protection circuit and KvmA sinterlock limitation and protection.

7.Adopt the digital circuit timer,grade according to R10 prioritycoefficient and can control time exactly

8.Examining bed is of electirc back rotation and is equipped with vibrating filter-ray device.

9.The photographic flat bed has a vibrating ray-filter.

10.The upright post has the bottom track only,which is easy whileinstallation.

TECHNICAL PARAMETER

ITEM

CONTENT

INDEX

Supply power

voltage

single phase 380V/220V ± 10%

 

frequency

50Hz ± 0.5Hz

 

capacity

10KVA

 

inner resistance

2O at 380Vl;0.6O at 220V

Max. Dc output voltage

90KV

100KV ?

Max. Dc output current

100mA

125mA ?

 

voltage

50~90KV

50~100KV ?

current

small focus 25mA and big focus 50 , 75 , 100 , 125mA four grades together

time

0.04~6.3s,23 grades together

 

capacity

7KVA

8.4KVA ?

 

model

XD4~2.9/100

focus

small focus 1.8mm big focus 4.3mm

Photographic bed(YZ-III)

Bed face(LXWXH)

1960mmX660X 700mm

Ray-filter

N28 Grid density N28

fo 100cm

Convergence distance fo 100cm

r8 Grid ratio R8

Max size of cassette for photograph

356mmX 432mm ( 14” X 17” )

Pillar of X-ray source unit

move lengthwise

(along the bottom track)

> 1350mm

move transverse

> 360mm

move up and down (along upright post)

720 -1800mm

MAX.CAPACITY OF PHOTOGRAY

Focus of X-ray tube

photograpy mA

Max.photograpy KV

Max.allowable expose time

small focus

25mA

90KV 100KV ?

1.0s

big focus

50 mA

90KV 95KV ?

6.3s

75 mA

90KV 95KV ?

5s

100 mA

85KV 90KV ?

3.2s

125mA

85KV ?

1.0s

 
 
Copyright © 2006 MARTVI Medical & Hospital Supply